Technical Tips

Master reset information on SI9000+ Enamel Rater gauge

In the event of a major change, error or fault it may become necessary to completely reset the instrument. This is always required when the operating mode is changed. A reset will erase all stored settings and reading from the unit's memory.

To perform a master reset proceed as follows;

A) Switch off power to the unit.
B) Simultaneously hold down the SET, up_arrowanddown_arrowkeys.
C) Switch on the power.
D) After 2 seconds release the keys and the system will be reset.

When a master reset is performed the operator has the opportunity to change the default times and voltage settings (factory set to 4 seconds and 6.3V) using the SET, .and .keys. The new default values are stored in battery backed memory and are retained until another master reset is performed. The rating voltage and duration can be checked during operation as described in Section 3.2 of the manual.

For further information please refer to the manual or contact the Technical Support Desk at Sencon.